Development of a microfluidic design for an automatic lab-on-chip operation
Puttaraksa, N., Whitlow, H. J., Napari, M., Meriläinen, L., & Gilbert, L. (2016). Development of a microfluidic design for an automatic lab-on-chip operation. Microfluidics and Nanofluidics, 20(142). https://doi.org/10.1007/s10404-016-1808-0
Published in
Microfluidics and NanofluidicsDate
2016Discipline
Solu- ja molekyylibiologiaFysiikkaNanoscience CenterKiihdytinlaboratorioCell and Molecular BiologyPhysicsNanoscience CenterAccelerator LaboratoryCopyright
© Springer-Verlag Berlin Heidelberg 2016. This is a final draft version of an article whose final and definitive form has been published by Springer. Published in this repository with the kind permission of the publisher.
Simple and easy to use are the keys for developing lab-on-chip technology. Here, a new microfluidic circuit has been designed for an automatic lab-on-chip operation (ALOCO) device. This chip used capillary forces for controlled and precise manipulation of liquids, which were loaded in sequence from different flowing directions towards the analysis area. Using the ALOCO design, a non-expert user is able to operate the chip by pipetting liquids into suitable inlet reservoirs. To test this design, microfluidic devices were fabricated using the programmable proximity aperture lithography technique. The operation of the ALOCO chip was characterized from the flow of red-, blue- and un-dyed deionized water. Experimental result indicated that red water, which filled first the analysis area, was substituted entirely with blue water. Controlled sequential flows of these water in the ALOCO device are demonstrated in this paper.
Publisher
SpringerISSN Search the Publication Forum
1613-4982Publication in research information system
https://converis.jyu.fi/converis/portal/detail/Publication/26248203
Metadata
Show full item recordCollections
Related items
Showing items with similar title or keywords.
-
Development of MeV ion beam lithography technique for microfluidic applications
Puttaraksa, Nitipon (University of Jyväskylä, 2011) -
Fabrication of microfluidic devices through deep wet etching
Öçal, Süha (2019)Soda-lime glass is a commonly used, cheap and accessible material. Just like any other glass, it offers unique optical properties. Most of the incoming light is transmitted through the glass, which makes soda-lime an ... -
Microfluidic cancer cell sorting using dielectrophoresis
Balogh, Péter (2020)Mikrofluidiikka mahdollistaa mikrobiologisten partikkeleiden, kuten solujen, viruksien, proteiinien ja DNA:n analysoinnin. Se pystyy täyttämään näiden tiukat elinolojen vaatimukset, joita ovat paine, lämpötila ja kemiallinen ... -
Crowdsourced analysis of fungal growth and branching on microfluidic platforms
Hopke, Alex; Mela, Alex; Ellett, Felix; Carter-House, Derreck; Peña, Jesús F.; Stajich, Jason E.; Altamirano, Sophie; Lovett, Brian; Egan, Martin; Kale, Shiv; Kronholm, Ilkka; Guerette, Paul; Szewczyk, Edyta; McCluskey, Kevin; Breslauer, David; Shah, Hiral; Coad, Bryan R.; Momany, Michelle; Irimia, Daniel (Public Library of Science (PLoS), 2021)Fungal hyphal growth and branching are essential traits that allow fungi to spread and proliferate in many environments. This sustained growth is essential for a myriad of applications in health, agriculture, and industry. ... -
MeV ion beam lithography of high aspect ratio structures with a focused or aperture-shaped beam for applications in biomedical studies and microfluidics
Gorelick, Sergey (University of Jyväskylä, 2008)