Robustness of radiation beam profile measurements
Tekijät
Päivämäärä
2023Tekijänoikeudet
© The Author(s)
In this master’s thesis, profile measurements of PTW Semiflex 3D and PTW Semiflex ionisation chambers in addition to IBA MatriXX matrix detector were compared to each other using different measurement set-ups. The ionisation chamber measure- ments were done in a water phantom in horizontal and vertical orientations. Using two different set-ups in the horizontal orientation. The first horizontal set-up used the current clinical quality assurance program in Tampere University Hospital with the wire of the chamber lying deep in the water phantom. The second horizontal measurements were done with more of the wire in the radiation field by having the wire closer to the surface of the water. Measurements were done with 6, 10 and 15 MV photon fields. Flattened fields were measured with 6 and 15 MV energies. Flattening filter free fields were measured for the 6 and 10 MV fields. Field size was set to 20x20 cm². All of the measurements were repeated with adding 5° wedge in the radiation beam.
Symmetry, flatness and width of left and right penumbras were compared across the measurements. The symmetry and flatness of the vertically orientated ionisation chamber were in agreement with the matrix detector measurements for all energies and for the open and wedge fields. The results were similar amongst the chamber type. Horizontally orientated Semiflex 3D deviated significantly on the x-axis from the other measurements in symmetry and flatness. The same effect was not seen on the y-axis. The amount of wire in the radiation field did not affect the measured profiles. Due to the observations that were made, orientation of the ionisation chamber was changed to vertical in the radiation therapy quality assurance program in the clinic.
...
Asiasanat
ionisation chamber matrix detector radiation therapy QA radiotherapy QA ionisoiva säteily säteily säteilyannokset mittausmenetelmät fotonit säteilyfysiikka sädehoito sironta hiukkaskiihdyttimet ionising radiation radiation radiation doses measuring methods photons radiation physics radiotherapy scattering (physics) particle accelerators
Metadata
Näytä kaikki kuvailutiedotKokoelmat
- Pro gradu -tutkielmat [29740]
Lisenssi
Samankaltainen aineisto
Näytetään aineistoja, joilla on samankaltainen nimeke tai asiasanat.
-
Ultra-high doses tests on 28nm CMOS technology for high energy physics application
Ciachera, Geoffrey (2024)This work is about the qualification of next generation of electronics detectors for high energy physics application. A new upgrade of the Large Hadron Collider, designated as High Luminosity Large Hadron Collider, is ... -
The pion single-event latch-up cross-section enhancement : mechanisms and consequences for accelerator hardness assurance
Coronetti, Andrea; Alia Garcia, Ruben; Cerutti, Francesco; Hajdas, Wojtek; Söderström, Daniel; Javanainen, Arto; Saigne, Frederic (Institute of Electrical and Electronics Engineers (IEEE), 2021)Pions make up a large part of the hadronic environment typical of accelerator mixed-fields. Characterizing device cross-sections against pions is usually disregarded in favour of tests with protons, whose single-event ... -
Electromagnetic wave propagation in non-homogeneous waveguides
Poretskii, Aleksandr (University of Jyväskylä, 2015)We investigate an electromagnetic waveguide, having several cylindrical ends. The waveguide is assumed to be empty and to have a perfectly conductive boundary. We study the electromagnetic field, excited in the waveguide ... -
Analysis of the photoneutron field near the THz dump of the CLEAR accelerator at CERN with SEU measurements and simulations
Lerner, Giuseppe; Coronetti, Andrea; Kempf, Jean Mael; Garcia Alia, Ruben; Cerutti, Francesco; Prelipcean, Daniel; Cecchetto, Matteo; Gilardi, Antonio; Farabolini, Wilfrid; Corsini, Roberto (Institute of Electrical and Electronics Engineers (IEEE), 2022)We study the radiation environment near the THz dump of the CLEAR electron accelerator at CERN, using FLUKA simulations and Single Event Upset measurements taken with 32 Mbit ISSI SRAM memories. The main focus is on the ... -
Mechanisms of Electron-Induced Single Event Latchup
Tali, Maris; Alía, Rubéen García; Brugger, Markus; Ferlet-Cavrois, Veronique; Corsini, Roberto; Farabolini, Wilfrid; Javanainen, Arto; Santin, Giovanni; Polo, Cesar Boatella; Virtanen, Ari (Institute of Electrical and Electronics Engineers, 2019)In this paper, possible mechanisms by which electrons can induce single-event latchups in electronics are discussed. The energy deposition and the nuclear fragments created by electrons in silicon are analyzed in this ...
Ellei toisin mainittu, julkisesti saatavilla olevia JYX-metatietoja (poislukien tiivistelmät) saa vapaasti uudelleenkäyttää CC0-lisenssillä.