Palosaari, M., Kinnunen, K., Julin, J., Laitinen, M., Napari, M., Sajavaara, T., Doriese, W. B., Fowler, J., Reintsema, C., Swetz, D., Schmidt, D., Ullom, J., & Maasilta, I. (2014). Transition-Edge Sensors for Particle Induced X-ray Emission Measurements. Journal of Low Temperature Physics, 176(3-4), 285-290. https://doi.org/10.1007/s10909-013-1004-5
Published in
Journal of Low Temperature PhysicsAuthors
Date
2014Copyright
© Springer Science+Business Media New York 2013. This is a final draft version of an article whose final and definitive form has been published by Springer. Published in this repository with the kind permission of the publisher.
In this paper we present a new measurement setup, where a transitionedge
sensor detector array is used to detect X-rays in particle induced X-ray emission
measurements with a 2 MeV proton beam. Transition-edge sensors offer orders
of magnitude improvement in energy resolution compared to conventional
silicon or germanium detectors, making it possible to recognize spectral lines
in materials analysis that have previously been impossible to resolve, and to get
chemical information from the elements. Our sensors are cooled to the operation
temperature (∼ 65 mK) with a cryogen-free adiabatic demagnetization refrigerator,
which houses a specially designed X-ray snout that has a vacuum tight window
to couple in the radiation. For the best pixel, the measured instrumental energy
resolution was 3.06 eV full width at half maximum at 5.9 keV. We discuss the current
status of the project, benefits of transition-edge sensors when used in particle
induced X-ray emission spectroscopy, and the results from the first measurements.
...
Publisher
Springer New York LLCISSN Search the Publication Forum
0022-2291Publication in research information system
https://converis.jyu.fi/converis/portal/detail/Publication/23756521
Metadata
Show full item recordCollections
Related items
Showing items with similar title or keywords.
-
Transition-edge sensors for particle-induced X-ray emission spectrometry
Käyhkö, Marko (University of Jyväskylä, 2018)A transition-edge sensor array has been applied to particle-induced X-ray emission. We have characterized the detector for quantitative analysis and obtained minimum detection limit over two orders of magnitude better ... -
Development and applications of transition-edge sensors
Palosaari, Mikko (University of Jyväskylä, 2015)This thesis focuses on the theory, development and applications of Transition- Edge Sensor (TES) arrays. The main results are the design and development of a fabrication process for a 256 pixel X-ray TES detector with a ... -
Characterizing a dc-SQUID and a Mo/TiW/Cu -trilayer transition-edge sensor array
Helenius, Ari (2017)Tutkielmassa määritettiin uuden suprajohtavan kvantti-interferenssilaitteen ja 256 pikseliä sisältävän sirun transitioreunailmaisimien ominaisuuksia. Komponentit vaativat erittäin matalia lämpötiloja ja suojauksen suurilta ... -
Transition-edge sensor array development
Palosaari, Mikko (2009) -
Simultaneous Noise and Impedance Fitting to Transition-Edge Sensor Data Using Differential Evolution
Helenius, A. P.; Puurtinen, T. A.; Kinnunen, K. M.; Maasilta, I. J. (Springer Science and Business Media LLC, 2020)We discuss a robust method to simultaneously fit a complex multi-body model both to the complex impedance and the noise data for transition-edge sensors. It is based on a differential evolution (DE) algorithm, providing ...