Determination of absolute internal conversion coefficients using the SAGE spectrometer
Sorri, J., Greenlees, P., Papadakis, P., Konki, J., Cox, D., Auranen, K., Partanen, J., Sandzelius, M., Pakarinen, J., Rahkila, P., Uusitalo, J., Herzberg, R.-D., Smallcombe, J., Davies, P.J., Barton, C.J., & Jenkins, D.G. (2016). Determination of absolute internal conversion coefficients using the SAGE spectrometer. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 812, 24-32. https://doi.org/10.1016/j.nima.2015.12.041
Julkaistu sarjassa
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated EquipmentTekijät
Päivämäärä
2016Tekijänoikeudet
© 2016 Elsevier B.V. This is a final draft version of an article whose final and definitive form has been published by Elsevier. Published in this repository with the kind permission of the publisher.
A non-reference based method to determine internal conversion coefficients using the SAGE spectrometer is carried out for transitions in the nuclei of 154Sm, 152Sm and 166Yb. The Normalised-Peak-to-Gamma method is in general an efficient tool to extract internal conversion coefficients. However, in many cases the required well-known reference transitions are not available. The data analysis steps required to determine absolute internal conversion coefficients with the SAGE spectrometer are presented. In addition, several background suppression methods are introduced and an example of how ancillary detectors can be used to select specific reaction products is given. The results obtained for ground-state band E2 transitions show that the absolute internal conversion coefficients can be extracted using the methods described with a reasonable accuracy. In some cases of less intense transitions only an upper limit for the internal conversion coefficient could be given.
Julkaisija
Elsevier B.V.ISSN Hae Julkaisufoorumista
0168-9002Asiasanat
Julkaisu tutkimustietojärjestelmässä
https://converis.jyu.fi/converis/portal/detail/Publication/25456031
Metadata
Näytä kaikki kuvailutiedotKokoelmat
Samankaltainen aineisto
Näytetään aineistoja, joilla on samankaltainen nimeke tai asiasanat.
-
Identification and reconstruction of low-energy electrons in the ProtoDUNE-SP detector
DUNE Collaboration (American Physical Society (APS), 2023)Measurements of electrons from νe interactions are crucial for the Deep Underground Neutrino Experiment (DUNE) neutrino oscillation program, as well as searches for physics beyond the standard model, supernova neutrino ... -
Development of a high energy resolution gas ionization detector for a recoil spectrometer
Julin, Jaakko (2011) -
Measurements of Low-Energy Protons using a Silicon Detector for Application to SEE Testing
Cazzaniga, Carlo; Alia, Ruben Garcia; Coronetti, Andrea; Bilko, Kacper; Morilla, Yolanda; Martin-Holgado, Pedro; Kastriotou, Maria; Frost, Christopher D. (IEEE, 2022)A silicon detector with a fast electronics chain is used for the dosimetry of protons in the range 0.5-5 MeV at the CNA 3 MV Tandem laboratory in Seville, Spain. In this configuration, measurements can be performed in ... -
Electron spectroscopy with the SAGE spectrometer
Sorri, Juha (University of Jyväskylä, 2016)A device called the SAGE (Silicon And GErmanium) spectrometer, designed for in-beam electron spectroscopy, is introduced in this work. A detailed description to the technical aspects of the device is given along with ... -
Conversion electron spectroscopy at IGISOL
Rinta-Antila, Sami; Trzaska, Wladyslaw; Rissanen, Juho (Springer Netherlands, 2014)Conversion elecron spectroscopy has been an important part of the nuclear spectrocopy research at the Department of Physics of the University of Jyv¨askyl¨a since the commissioning of the first cyclotron in the mid 1970s. ...
Ellei toisin mainittu, julkisesti saatavilla olevia JYX-metatietoja (poislukien tiivistelmät) saa vapaasti uudelleenkäyttää CC0-lisenssillä.