Validity of the Wrist-Worn Polar Vantage V2 to Measure Heart Rate and Heart Rate Variability at Rest
Nuuttila, O.-P., Korhonen, E., Laukkanen, J., & Kyröläinen, H. (2022). Validity of the Wrist-Worn Polar Vantage V2 to Measure Heart Rate and Heart Rate Variability at Rest. Sensors, 22(1), Article 137. https://doi.org/10.3390/s22010137
Published in
SensorsDate
2022Copyright
© 2021 by the authors. Licensee MDPI, Basel, Switzerland
Heart rate (HR) and heart rate variability (HRV) can be monitored with wearable devices throughout the day. Resting HRV in particular, reflecting cardiac parasympathetic activity, has been proposed to be a useful marker in the monitoring of health and recovery from training. This study examined the validity of the wrist-based photoplethysmography (PPG) method to measure HR and HRV at rest. Recreationally endurance-trained participants recorded pulse-to-pulse (PP) and RR intervals simultaneously with a PPG-based watch and reference heart rate sensor (HRS) at a laboratory in a supine position (n = 39; 5-min recording) and at home during sleep (n = 29; 4-h recording). In addition, analyses were performed from pooled laboratory data (n = 11340 PP and RR intervals). Differences and correlations were analyzed between the HRS- and PPG-derived HR and LnRMSSD (the natural logarithm of the root mean square of successive differences). A very good agreement was found between pooled PP and RR intervals with a mean bias of 0.17 ms and a correlation coefficient of 0.993 (p < 0.001). In the laboratory, HR did not differ between the devices (mean bias 0.0 bpm), but PPG slightly underestimated the nocturnal recordings (bias −0.7 bpm, p < 0.001). PPG overestimated LnRMSSD both in the laboratory (bias 0.20 ms, p < 0.001) and nocturnal recordings (bias 0.17 ms, p < 0.001). However, very strong intraclass correlations in the nocturnal recordings were found between the devices (HR: 0.998, p < 0.001; LnRMSSD: 0.931, p < 0.001). In conclusion, PPG was able to measure HR and HRV with adequate accuracy in recreational athletes. However, when strict absolute values are of importance, systematic overestimation, which seemed to especially concern participants with low LnRMSSD, should be acknowledged.
...
Publisher
MDPI AGISSN Search the Publication Forum
1424-8220Keywords
Dataset(s) related to the publication
Nuuttila, Olli-Pekka. (2023). Monitoring recovery and training responses from different types of endurance exercises and training protocols in recreational runners: implications for individual training prescription . University of Jyväskylä. https://doi.org/10.17011/jyx/dataset/84680. https://urn.fi/URN:NBN:fi:jyu-202301031038Publication in research information system
https://converis.jyu.fi/converis/portal/detail/Publication/103517975
Metadata
Show full item recordCollections
- Liikuntatieteiden tiedekunta [3164]
Additional information about funding
This research was partially funded by Polar Electro Oy.License
Related items
Showing items with similar title or keywords.
-
Validity of three smartwatches in estimating energy expenditure during outdoor walking and running
Le, Shenglong; Wang, Xiuqiang; Zhang, Tao; Lei, Si Man; Cheng, Sulin; Yao, Wu; Schumann, Moritz (Frontiers Media SA, 2022)Commercially wrist-worn devices often present inaccurate estimations of energy expenditure (EE), with large between-device differences. We aimed to assess the validity of the Apple Watch Series 6 (AW), Garmin FENIX 6 (GF) ... -
Contextualizing Everyday Data Literacies : The Case of Recreational Runners
Palsa, Lauri; Mertala, Pekka (Informa UK Limited, 2023)Data literacy is typically described in a decontextualized manner, and many data literacy frameworks are detached from the “messy” realities of everyday life. In the present study, we selected a specific context (recreational ... -
Thermal Measurement of Arterial Pulse using Heat Flux Sensors
Immonen, Antti; Pettersson, Ante B. V.; Levikari, Saku; Peltonen, Heikki; Kyröläinen, Heikki; Silventoinen, Pertti; Kuisma, Mikko (Institute of Electrical and Electronics Engineers (IEEE), 2024)In this paper, the novel concept of using heat flux sensors (HFS) to measure arterial pulse on the skin surface is validated. The heat flux (HF) signal is compared with simultaneously measured electrocardiogram (ECG) and ... -
Puettava teknologia valmennuksen tukena
Löppönen, Antti; Rantalainen, Timo (Liikuntatieteellinen seura, 2022)Sensoriteknologia on jo melko laajasti käytössä valmennuksessa. Lopullinen läpimurto on kuitenkin vasta tulossa. -
Validity of IMU sensors for assessing features of walking in laboratory and outdoor environments among older adults
Matikainen-Tervola, Emmi; Cronin, Neil; Aartolahti, Eeva; Sihvonen, Sanna; Sansgiri, Sailee; Finni, Taija; Mattila, Olli-Pekka; Rantakokko, Merja (Elsevier, 2024)Introduction: IMU sensors (three-dimensional accelerometer, gyroscope and magnetometer) enable assessment of walking in older adults outside the laboratory. We studied whether IMUs are valid for detecting walking parameters ...