Spectroscopic method to study low charge state ion and cold electron population in ECRIS plasma
Kronholm, R., Kalvas, T., Koivisto, H., & Tarvainen, O. (2018). Spectroscopic method to study low charge state ion and cold electron population in ECRIS plasma. Review of Scientific Instruments, 89(4), Article 043506. https://doi.org/10.1063/1.5023434
Julkaistu sarjassa
Review of Scientific InstrumentsPäivämäärä
2018Tekijänoikeudet
© AIP Publishing, 2018. Published in this repository with the kind permission of the publisher.
The results of optical emission spectroscopy experiments probing the cold electron population of a 14 GHz Electron Cyclotron Resonance Ion Source (ECRIS) are reported. The study has been conducted with a high resolution spectrometer and data acquisition setup developed specifically for the diagnostics of weak emission line characteristic to ECRIS plasmas. The optical emission lines of low charge state ions and neutral atoms of neon have been measured and analyzed with the line-ratio method. The aforementioned electron population temperature of the cold electron population (Te < 100 eV) is determined for Maxwell-Boltzmann and Druyvesteyn energy distributions to demonstrate the applicability of the method. The temperature was found to change significantly when the extraction voltage of the ion source is turned on/off. In the case of the Maxwellian distribution, the temperature of the cold electron population is 20 ± 10 eV when the extraction voltage is off and 40 ± 10 eV when it is on. The optical emission measurements revealed that the extraction voltage also affects both neutral and ion densities. Based on the rate coefficient analysis with the aforementioned temperatures, switching the extraction voltage off decreases the rate coefficient of neutral to 1+ ionization to 42% and 1+ to 2+ ionization to 24% of the original. This suggests that switching the extraction voltage on favors ionization to charge states ≥2+ and, thus, the charge state distributions of ECRIS plasmas are probably different with the extraction voltage on/off. It is therefore concluded that diagnostics results of ECRIS plasmas obtained without the extraction voltage are not depicting the plasma conditions in normal ECRIS operation.
...
Julkaisija
AIP Publishing LLCISSN Hae Julkaisufoorumista
0034-6748Asiasanat
Julkaisu tutkimustietojärjestelmässä
https://converis.jyu.fi/converis/portal/detail/Publication/27992628
Metadata
Näytä kaikki kuvailutiedotKokoelmat
Rahoittaja(t)
Euroopan komissioRahoitusohjelmat(t)
Research infrastructures, H2020
The content of the publication reflects only the author’s view. The funder is not responsible for any use that may be made of the information it contains.
Lisätietoja rahoituksesta
The authors would like to thank J. Komppula, M. Liimatainen, and the JYFL electrical and mechanical workshop for their help in measurement technical issues. This project has received funding from European Union’s Horizon 2020 research and innovation programme under Grant Agreement No. 654002 and the Academy of Finland under the Finnish Centre of Excellence Programme 2012–2017 (Nuclear and Accelerator Based Physics Research at JYFL). ...Samankaltainen aineisto
Näytetään aineistoja, joilla on samankaltainen nimeke tai asiasanat.
-
Influence of axial mirror ratios on the kinetic instability threshold in electron cyclotron resonance ion source plasma
Toivanen, V.; Bhaskar, B. S.; Koivisto, H.; Maunoury, L.; Tarvainen, O.; Thuillier, T. (American Institute of Physics, 2022)Electron Cyclotron Resonance (ECR) ion source plasmas are prone to kinetic instabilities. The onset of the instabilities manifests as emission of microwaves, bursts of electrons expelled from the plasma volume, and the ... -
Slit extraction and emittance results of a permanent magnet minimum-B quadrupole electron cyclotron resonance ion source
Kosonen, Sami; Kalvas, Taneli; Koivisto, Hannu; Tarvainen, Olli; Toivanen, Ville (Elsevier, 2024)We present an experimental and simulation study of high charge state ion beams produced with a permanent magnet electron cyclotron resonance ion source (ECRIS) with minimum-B quadrupole magnetic field topology and slit ... -
Bremsstrahlung Emission of an X-Band Permanent Magnet Minimum-B Quadrupole Electron Cyclotron Resonance Ion Source
Tarvainen, Olli; Toivanen, Ville; Kalvas, Taneli; Koivisto, Hannu A.; Kosonen, Sami (Institute of Electrical and Electronics Engineers (IEEE), 2024)We have carried out bremsstrahlung measurements on a permanent magnet minimum-B quadrupole electron cyclotron resonance (ECR) ion source operating at 10.3–11.5 GHz microwave frequencies. The bremsstrahlung spectral temperature ... -
Study of gasdynamic electron cyclotron resonance plasma vacuum ultraviolet emission to optimize negative hydrogen ion production efficiency
Lapin, R. L.; Skalyga, V. A.; Izotov, I.; Razin, S. V.; Shaposhnikov, R. A.; Vybin, S. S.; Bokhanov, A. F.; Kazakov, M. Yu.; Tarvainen, O. (American Institute of Physics, 2020)Negative hydrogen ion sources are used as injectors into accelerators and drive the neutral beam heating in ITER. Certain processes in low-temperature hydrogen plasmas are accompanied by the emission of vacuum ultraviolet ... -
Characteristic Kα emission of electron cyclotron resonance ion source plasmas
Sakieldien, Moenir (University of Jyväskylä, 2018)This thesis presents the results of an investigation of the volumetric Kα emission rate/inner shell ionization rate from an Electron Cyclotron Resonance Ion Source (ECRIS) plasma tuned predominantly for high charge state ...
Ellei toisin mainittu, julkisesti saatavilla olevia JYX-metatietoja (poislukien tiivistelmät) saa vapaasti uudelleenkäyttää CC0-lisenssillä.