Näytä suppeat kuvailutiedot

dc.contributor.authorRekonen, Ahti
dc.date.accessioned2024-01-15T12:42:05Z
dc.date.available2024-01-15T12:42:05Z
dc.date.issued1972
dc.identifier.isbn978-951-39-9994-0
dc.identifier.urihttps://jyx.jyu.fi/handle/123456789/92815
dc.description.abstractThe present work introduces a new transversal tomography procedure, where the radiation source is a gamma-rays emitting radionuclide and the detector a scintillation counter. The procedure involves transversal and rotary motion of the collimated gamma-ray beam at the emitter-detector unit relative to the subject. The modulation transfer functions are calculated for each component unit, for the motions and finally for the complete system. A comparison indicates that the developed transversal transmission gammatomography (TTGT) ranks in terms of image resolution above radioisotope scintigraphy though lower than roentgen transversal tomography. The radiation load in the TTGT process is clearly smaller than that of conventional X-ray or scintigraphic methods. Some results and photographs of phantom and patient tomograms are presented.en
dc.language.isoeng
dc.rightsIn Copyright
dc.titleA method for transversal tomography using gamma radiation : transversal transmission gammatomography
dc.typedoctoral thesis
dc.identifier.urnURN:ISBN:978-951-39-9994-0
dc.contributor.yliopistoUniversity of Jyväskyläen
dc.contributor.yliopistoJyväskylän yliopistofi
dc.type.coarhttp://purl.org/coar/resource_type/c_db06
dc.rights.accesslevelopenAccess
dc.type.publicationdoctoralThesis
dc.rights.urlhttps://rightsstatements.org/page/InC/1.0/
dc.date.digitised2014


Aineistoon kuuluvat tiedostot

Thumbnail

Aineisto kuuluu seuraaviin kokoelmiin

Näytä suppeat kuvailutiedot

In Copyright
Ellei muuten mainita, aineiston lisenssi on In Copyright