Reliability of upper-limb diaphyseal mineral and soft-tissue measurements using peripheral Quantitative Computed Tomography (pQCT)
Jenkins, M. A., Hart, N. H., Rantalainen, T., Chivers, P., Newton, R. U., & Nimphius, S. (2018). Reliability of upper-limb diaphyseal mineral and soft-tissue measurements using peripheral Quantitative Computed Tomography (pQCT). Journal of Musculoskeletal and Neuronal Interactions, 18(4), 438-445. http://www.ismni.org/jmni/pdf/74/jmni_18_438.pdf
Julkaistu sarjassa
Journal of Musculoskeletal and Neuronal InteractionsTekijät
Päivämäärä
2018Oppiaine
Gerontologian tutkimuskeskusHyvinvoinnin tutkimuksen yhteisöGerontology Research CenterSchool of WellbeingTekijänoikeudet
© The Authors, 2018.
Objectives: To quantify between-day reliability of upper-body diaphyseal measurements (radius, ulna, humerus) using
peripheral Quantitative Computed Tomography (pQCT). - Methods: Fourteen males (age: 25.8±2.3 years,) underwent
repeat pQCT scans (one to two days apart) at mid-shaft ulna (60%), mid-shaft radius (60%) and mid-shaft humerus
(50%) cross-sections of the non-dominant limb. Intraclass correlation coefficients (ICC) and coefficients of variation (CV)
were determined for musculoskeletal morphology variables. - Results: Reliability was excellent (ICC: 0.76–0.99; CV: 1.3–
7.3) at all sites for bone mass, stress-strain index, endocortical and pericortical radius, endocortical volumetric bone
mineral density (vBMD), muscle area, total area, non-cortical area, and cortical area. Reliability was good to excellent
(ICC: 0.58–0.80; CV: 0.6–3.7) for polar vBMD and mid-cortical vBMD; fair to excellent (ICC: 0.30–0.88; CV: 0.5–8.0)
for muscle density and cortical density; and fair to good (ICC: 0.25–0.60; CV: 3.4–7.6) for pericortical vBMD. Average
reliability across the three sites was excellent (ICC ≥0.77; CV ≤8.0). - Conclusions: Overall between-day reliability of pQCT
was excellent for the mid-shaft ulna, radius and humerus. pQCT provides a reliable and feasible body composition and
skeletal morphology assessment tool for upper limb longitudinal investigations in scientific and clinic settings.
...
Julkaisija
Hylonome PublicationsISSN Hae Julkaisufoorumista
1108-7161Asiasanat
Alkuperäislähde
http://www.ismni.org/jmni/pdf/74/jmni_18_438.pdfJulkaisu tutkimustietojärjestelmässä
https://converis.jyu.fi/converis/portal/detail/Publication/28836495
Metadata
Näytä kaikki kuvailutiedotKokoelmat
- Liikuntatieteiden tiedekunta [3123]
Lisenssi
Samankaltainen aineisto
Näytetään aineistoja, joilla on samankaltainen nimeke tai asiasanat.
-
Feasibility of acousto-electric tomography
Jensen, Bjørn; Kirkeby, Adrian; Knudsen, Kim (IOP Publishing, 2024)In acousto-electric tomography (AET) the goal is to reconstruct the electric conductivity in a domain from electrostatic boundary measurements of corresponding currents and voltages, while the domain is perturbed by a ... -
Identifying and Assessing Inter-Muscular Fat at the Distal Diaphyseal Femur Measured by Peripheral Quantitative Computed Tomography (pQCT)
Owen, Patrick J.; Hart, Nicolas H.; Latella, Christopher; Hendy, Ashlee M.; Lamon, Séverine; Rantalainen, Timo (International Society for Clinical Densitometry; Elsevier Inc., 2021)INTRODUCTION Inter/intramuscular fat can be assessed with peripheral Quantitative Computed Tomography (pQCT) and is of interest as an indicator of ‘muscle quality’. Typical pQCT scan sites (forearm, lower leg) have a low ... -
Low-Dose Mouse Micro-CT Image Segmentation Based on Multi-Resolution Multi-Organ Shape Prior Knowledge Model
Chen, Zhonghua; Wang, Hongkai; Cong, Fengyu; Kettunen, Lauri (IEEE, 2022)Automatic segmentation of computed tomography (CT) images of mice is a step toward computer-assisted preclinical image analysis. Due to the low image quality of micro-CT images, fully-automatic methods may not achieve ... -
Optimal recovery of a radiating source with multiple frequencies along one line
Brander, Tommi; Ilmavirta, Joonas; Piiroinen, Petteri; Tyni, Teemu (American Institute of Mathematical Sciences (AIMS), 2020)We study an inverse problem where an unknown radiating source is observed with collimated detectors along a single line and the medium has a known attenuation. The research is motivated by applications in SPECT and beam ... -
Novel Methodology for Localizing and Studying Insect Dorsal Rim Area Morphology in 2D and 3D
Jie, Vun Wen; Miettinen, Arttu; Baird, Emily (MDPI, 2023)Polarized light-based navigation in insects is facilitated by a polarization-sensitive part of the eye, the dorsal rim area (DRA). Existing methods to study the anatomy of the DRA are destructive and time-consuming. We ...
Ellei toisin mainittu, julkisesti saatavilla olevia JYX-metatietoja (poislukien tiivistelmät) saa vapaasti uudelleenkäyttää CC0-lisenssillä.