The effect of cavity tuning on oxygen beam currents of an A-ECR type 14 GHz electron cyclotron resonance ion source
Tarvainen, O., Orpana, J., Kronholm, R., Kalvas, T., Laulainen, J., Koivisto, H., Izotov, I., Skalyga, V., & Toivanen, V. (2016). The effect of cavity tuning on oxygen beam currents of an A-ECR type 14 GHz electron cyclotron resonance ion source. Review of Scientific Instruments, 87(9), Article 093301. https://doi.org/10.1063/1.4962026
Julkaistu sarjassa
Review of Scientific InstrumentsTekijät
Päivämäärä
2016Tekijänoikeudet
© Published by AIP Publishing.
The efficiency of the microwave-plasma coupling plays a significant role in the production of highly
charged ion beams with electron cyclotron resonance ion sources (ECRISs). The coupling properties
are affected by the mechanical design of the ion source plasma chamber and microwave launching
system, as well as damping of the microwave electric field by the plasma. Several experiments
attempting to optimize the microwave-plasma coupling characteristics by fine-tuning the frequency of
the injected microwaves have been conducted with varying degrees of success. The inherent difficulty
in interpretation of the frequency tuning results is that the effects of microwave coupling system and
the cavity behavior of the plasma chamber cannot be separated. A preferable approach to study the
effect of the cavity properties of the plasma chamber on extracted beam currents is to adjust the
cavity dimensions. The results of such cavity tuning experiments conducted with the JYFL 14 GHz
ECRIS are reported here. The cavity properties were adjusted by inserting a conducting tuner rod
axially into the plasma chamber. The extracted beam currents of oxygen charge states O3+–O7+
were recorded at various tuner positions and frequencies in the range of 14.00–14.15 GHz. It was
observed that the tuner position affects the beam currents of high charge state ions up to several
tens of percent. In particular, it was found that at some tuner position / frequency combinations
the plasma exhibited “mode-hopping” between two operating regimes. The results improve the
understanding of the role of plasma chamber cavity properties on ECRIS performances.
...
Julkaisija
American Institute of PhysicsISSN Hae Julkaisufoorumista
0034-6748Julkaisu tutkimustietojärjestelmässä
https://converis.jyu.fi/converis/portal/detail/Publication/26213577
Metadata
Näytä kaikki kuvailutiedotKokoelmat
Samankaltainen aineisto
Näytetään aineistoja, joilla on samankaltainen nimeke tai asiasanat.
-
Effect of double frequency heating on the lead afterglow beam currents of an electron cyclotron resonance ion source
Toivanen, V.; Bellodi, G.; Küchler, D.; Wenander, F.; Tarvainen, Olli (American Physical Society, 2017)The effect of double frequency heating on the performance of the CERN GTS-LHC 14.5 GHz Electron Cyclotron Resonance (ECR) ion source in afterglow mode is reported. The source of the secondary microwave frequency was operated ... -
Plasma response to amplitude modulation of the microwave power on a 14 GHz electron cyclotron resonance ion source
Neben, Derek; Tarvainen, Olli; Kronholm, Risto; Koivisto, Hannu; Kalvas, Taneli; Machicoane, Guillaume; Leitner, Daniela (AIP Publishing, 2018)This paper reports the effects of sinusoidal microwave power Amplitude Modulation (AM) on the performance of Electron Cyclotron Resonance (ECR) ion sources. The study was conducted on the 14 GHz ECR ion source ECR2 at the ... -
Periodic Beam Current Oscillations Driven by Electron Cyclotron Instabilities in ECRIS Plasmas
Tarvainen, Olli; Kalvas, Taneli; Koivisto, Hannu; Komppula, Jani; Kronholm, Risto; Laulainen, Janne; Izotov, I.; Mansfeld, D.; Skalyga, V.; Toivanen, V. (Institute of Applied Physics of the Russian Academy of Sciences, 2014)Experimental observation of cyclotron instabilities in electron cyclotron resonance ion source plasma operated in cwmode is reported. The instabilities are associated with strong microwave emission and a burst of energetic ... -
The effect of microwave power on the Ar9+ and Ar13+ optical emission intensities and ion beam currents in ECRIS
Kronholm, Risto; Sakildien, M.; Neben, D.; Koivisto, Hannu; Kalvas, Taneli; Tarvainen, Olli; Laulainen, J.; Jones, P. (AIP Publishing, 2018)The production of Ar9+ and Ar13+ ions in an ECRIS plasma and the efficiency of the ion beam extraction and transport of the resulting Ar9+ and Ar13+ ion beams have been studied with the JYFL 14 GHz ECRIS by using optical ... -
Influence of axial mirror ratios on the kinetic instability threshold in electron cyclotron resonance ion source plasma
Toivanen, V.; Bhaskar, B. S.; Koivisto, H.; Maunoury, L.; Tarvainen, O.; Thuillier, T. (American Institute of Physics, 2022)Electron Cyclotron Resonance (ECR) ion source plasmas are prone to kinetic instabilities. The onset of the instabilities manifests as emission of microwaves, bursts of electrons expelled from the plasma volume, and the ...
Ellei toisin mainittu, julkisesti saatavilla olevia JYX-metatietoja (poislukien tiivistelmät) saa vapaasti uudelleenkäyttää CC0-lisenssillä.